Refrigerated products moving on domestic or cross-border routes may pass through vehicles, cold rooms and receiving points before delivery. MaxiLog PDF is a pre-programmed, single-use temperature data logger that records shipment temperatures so the record can be checked at the end of each route.
It measures from -30°C to +70°C and records every 10 minutes for up to 90 days, with capacity for 14,400 data points. Its compact 94 mm × 20 mm × 7 mm format fits inside small packages, and the standard 30-minute start delay allows the logger to adjust to the cargo’s internal temperature before recording begins.
After delivery, connect the built-in USB-A interface to a computer to access PDF and CSV reports. The PDF includes a graph and summary, while the green and red LEDs provide a quick indication of operating status and whether a programmed alarm has been triggered.
- Up to 90 days of recording at 10-minute intervals for domestic and cross-border shipment routes.
- Automatic PDF and CSV report access through the built-in USB-A interface.
- Single-use, pre-programmed operation with a standard 30-minute start delay.
- Compact 94 mm × 20 mm × 7 mm format for small shipment packages.
- Programmable high and low alarm limits with green operating and red alarm LED indications.
- Push-button Start/Mark/Stop operation with up to 10 report marks for route events or handovers.
- Model: 7050-G2 MaxiLog PDF V3 Temperature Data Logger
- Recording Type: Single Use
- Temperature Measurement Range: -30°C to +70°C
- Temperature Accuracy: ±0.3°C
- Temperature Resolution: 0.1°C
- Data Storage Capacity: 14K data points (14,400)
- Battery Life / Type: CR2032 lithium coin cell, 18-month shelf life
- Data Sampling Interval: 10 minutes
- Start-Up Delay: 30 minutes
- Recording Duration: 90 days
- Alarm Programming: Programmable high and low limits; alarm triggered when limits are exceeded
- Alarm Modes: Exceed Limit
- Alarm Reset Options: Single Use: No reset option
- Alarm Notifications: Red LED flashes when alarm has been triggered
- Indicators: Green LED for on/recording; red LED for alarm
- Start-Up Options: Push button: Start/Mark/Stop
- Dimensions: 94 mm (H) × 20 mm (W) × 7 mm (D)
- Weight: 10.6 g / 0.37 oz.
- Quality Assurance Certifications: CE, UKCA, RTCA, NIST Traceable Certificate of Calibration, IP67 and RoHS compliant
- Minimum System Requirements: Windows XP, Vista, 7, 8 and 10; Mac and Android
- Report Formats: PDF and CSV
- Required Software: PDF Reader
- Computer Interface: USB A, built-in interface
- Warranty: 1 year
TempU Product Brochure - Humi Pak Malaysia: Download
Important: Do not open the sleeve bag until the trip is finished and you are ready to download the report.
1. Start the logger: Press and hold the button for at least 3 seconds until the OK LED flashes 5 times. The OK LED then flashes every 2 seconds to indicate that the logger is operating. The standard 30-minute start delay runs before temperature recording begins.
2. Set a mark if needed: Double-click the button. The OK LED flashes 3 times to confirm the mark. Up to 10 event marks can be added, and each mark appears on the graph at the corresponding point in time.
3. Stop the logger: Press and hold the button for at least 3 seconds until the Alarm LED flashes 5 times. Alternatively, connect the USB interface to a computer to stop the logger automatically.
4. Get the report: Connect the USB interface to a computer. You may need to wait until both LEDs become solid. The logger opens as a flash drive containing the data report in PDF and CSV formats.
MaxiLog PDF can be used to monitor temperature-sensitive shipments including:
- Pharmaceutical products
- Vaccines and biologics
- Life-science shipments
- Refrigerated food products
- Temperature-sensitive chemicals
- Domestic and cross-border cold-chain distribution
Is MaxiLog PDF suitable for domestic and cross-border cold-chain routes?
Yes, where shipment temperatures remain within its -30°C to +70°C measurement range and the journey fits within the 90-day recording duration.
Can the logger be reused for recurring deliveries?
No. MaxiLog PDF is single use, so a new logger is required for each shipment.
Does MaxiLog PDF require special software?
No dedicated logger software is required to retrieve the report. A PDF Reader is required to view the PDF file.
How do I know if a programmed alarm has been triggered?
The red LED flashes when an alarm condition has been triggered. The report can then be reviewed for the recorded temperature history.
How much data can MaxiLog PDF store?
It can store up to 14,400 data points and records at a fixed 10-minute sampling interval.
When should the sleeve bag be opened?
Only after the trip is finished and you are ready to download the report, in line with the manufacturer’s instructions.